What does mV/V mean on a load cell?
It is the bridge output at rated capacity per volt of excitation. A 2 mV/V cell excited at 5 V gives 10 mV at full load and half of that at half load.
From the mV/V rating and excitation, this load cell calculator works out the output voltage, then what an ADC such as the HX711 makes of it: counts per kilogram and grams per count. Enter the ADC’s input noise to see the resolution you can actually hold.
In the load unit above.
From the cell’s datasheet or calibration sheet.
Measure it at the cell (E+ to E−) if you can.
Above the tare, in the load unit above.
Optional: platform and fixture weight sitting on the cell.
Leave blank when the excitation is the reference. HX711: AVDD.
From the ADC datasheet. HX711 and ADS1232 use ±0.5·Vref / gain.
Type 50n for 50 nV. HX711 at gain 128: 50 nV at 10 SPS, 90 nV at 80 SPS (typical).
Ideal noise falls by √N; the settling time grows.
Bridge output at this load
5 mV
Noise sets the resolution, not the count size: the last 7.1 bits of each reading are noise.
Ratiometric: the ADC reference is the excitation, so counts do not change when the supply drifts.
Expected readings across the range
| % of capacity | Load (kg) | Bridge output | Net counts |
|---|---|---|---|
| 0 % | 0 | 0 V | 0 |
| 25 % | 2.5 | 2.5 mV | 1,073,742 |
| 50 % | 5 | 5 mV | 2,147,484 |
| 75 % | 7.5 | 7.5 mV | 3,221,225 |
| 100 % | 10 | 10 mV | 4,294,967 |
Output = rated output × excitation × load ÷ capacity, so a 2 mV/V cell at 5 V gives 10 mV at full load. On an HX711 at gain 128, the input range is ±19.53 mV across 2²⁴ codes, one code is 2.33 nV, and a 10 kg cell reads 429,497 counts per kg. Input noise, not count size, sets the limit: about 330 mg noise-free.
A load cell is a full strain gauge bridge calibrated at the factory. Its rating already includes the gauge factor and the mechanical design, so the load cell equation starts from mV/V.
A bipolar N-bit ADC spreads 2ᴺ codes over its input range. The HX711 and ADS1232 both specify a full scale of ±0.5 × Vref ÷ gain, and on those parts Vref is normally the excitation:
For noise, use the ADC’s input-referred RMS figure at your gain and data rate. Peak-to-peak noise is about 6.6 × RMS, because ±3.3 standard deviations hold 99.9 % of Gaussian samples. TI’s ADS1232 measurements land close to that, with a ratio of 6.5 at gain 128 and 10 SPS.
2 mV/V × 5 V = 10 mV at 10 kg, or 1 mV per kg. At gain 128 and AVDD = 5 V, the input range is ±19.53 mV, so the cell uses 51.2 % of it. One code is 2.33 nV, giving 429,497 counts per kg and 2.33 mg per count. With 50 nV RMS at 10 SPS, peak-to-peak noise is 330 nV, so the resolution you can hold is 330 mg, about 30,303 steps across the capacity (14.9 bits).
A 200 kg, 2 mV/V cell at 5 V has the same 10 mV full scale but only 50 µV per kg. The ADS1232 at gain 128 and 10 SPS has 17 nV RMS, so peak-to-peak noise is 112 nV and the resolution is 2.24 g, or 89,127 steps. TI measures about 85,000 counts for this same 10 mV signal over 60 seconds, within 5 % of the estimate.
A 5 kg, 1 mV/V cell at 5 V on an HX711 at gain 128 gives 429,497 counts per kg. A net reading of 1,000,000 counts above the tare is therefore 2.3283 kg. Treat that as a starting point: calibrate with a known weight, because the rated output of both cells cited on this page is only specified to ±10 %.
Bridge output at rated capacity. Compare it with the ADC input range before you pick a gain. For a rating not listed here, the tool at the top of the page doubles as an mV/V calculator.
| Rated output | 3.3 V | 5 V | 10 V |
|---|---|---|---|
| 1 mV/V | 3.3 mV | 5 mV | 10 mV |
| 1.5 mV/V | 4.95 mV | 7.5 mV | 15 mV |
| 2 mV/V | 6.6 mV | 10 mV | 20 mV |
| 3 mV/V | 9.9 mV | 15 mV | 30 mV |
Gain 128, AVDD and excitation 5 V, 24-bit, single readings. Noise is the datasheet’s typical input noise: 50 nV RMS at 10 SPS and 90 nV RMS at 80 SPS. For another cell or ADC, use the load cell resolution calculator above.
| Capacity | Rated output | Full scale | Counts per kg | Per count | Noise-free, 10 SPS | Noise-free, 80 SPS |
|---|---|---|---|---|---|---|
| 1 kg | 1 mV/V | 5 mV | 2,147,484 | 466 µg | 66 mg | 119 mg |
| 5 kg | 1 mV/V | 5 mV | 429,497 | 2.33 mg | 330 mg | 594 mg |
| 10 kg | 1 mV/V | 5 mV | 214,748 | 4.66 mg | 660 mg | 1.19 g |
| 20 kg | 1 mV/V | 5 mV | 107,374 | 9.31 mg | 1.32 g | 2.38 g |
| 10 kg | 2 mV/V | 10 mV | 429,497 | 2.33 mg | 330 mg | 594 mg |
| 50 kg | 2 mV/V | 10 mV | 85,899 | 11.6 mg | 1.65 g | 2.97 g |
| 100 kg | 2 mV/V | 10 mV | 42,950 | 23.3 mg | 3.3 g | 5.94 g |
| 200 kg | 2 mV/V | 10 mV | 21,475 | 46.6 mg | 6.6 g | 11.9 g |
Typical datasheet values. The HX711 datasheet gives noise only at gain 128, so peak-to-peak is 6.6 × RMS there; the ADS1232 peak-to-peak figures are TI’s measurements.
| ADC | Gain | Data rate | Input range | One code | RMS noise | Peak to peak | Noise-free bits |
|---|---|---|---|---|---|---|---|
| HX711, channel A | 128 | 10 SPS | ±19.53 mV | 2.33 nV | 50 nV | 330 nV | 16.9 |
| HX711, channel A | 128 | 80 SPS | ±19.53 mV | 2.33 nV | 90 nV | 594 nV | 16 |
| HX711, channel A | 64 | any | ±39.06 mV | 4.66 nV | not specified | not specified | not specified |
| HX711, channel B | 32 | any | ±78.13 mV | 9.31 nV | not specified | not specified | not specified |
| ADS1232 | 128 | 10 SPS | ±19.53 mV | 2.33 nV | 17 nV | 110 nV | 18.4 |
| ADS1232 | 128 | 80 SPS | ±19.53 mV | 2.33 nV | 44 nV | 247 nV | 17.3 |
| ADS1232 | 64 | 10 SPS | ±39.06 mV | 4.66 nV | 19 nV | 125 nV | 19.3 |
| ADS1232 | 64 | 80 SPS | ±39.06 mV | 4.66 nV | 48 nV | 307 nV | 18 |
It is the bridge output at rated capacity per volt of excitation. A 2 mV/V cell excited at 5 V gives 10 mV at full load and half of that at half load.
Vout = rated output × excitation × load ÷ capacity. A 10 kg, 2 mV/V cell at 5 V with 5 kg on it gives 5 mV.
Sensitivity = rated output × excitation ÷ capacity, which is the output per unit of load. A 10 kg, 2 mV/V cell at 5 V gives 10 mV ÷ 10 kg = 1 mV per kg.
Divide the bridge output by the excitation to get mV/V, then divide by the rated output and multiply by the capacity. On a 10 kg, 2 mV/V cell, 5 mV at 5 V is 1 mV/V, which is 5 kg. Calibrate with a known weight for the final scale factor.
±0.5 × AVDD ÷ gain. At 5 V that is ±19.53 mV at gain 128, ±39.06 mV at gain 64 and ±78.1 mV on channel B at gain 32. The datasheet’s text rounds these to ±20, ±40 and ±80 mV.
With a 10 kg, 2 mV/V cell at gain 128 and AVDD as the excitation, 429,497 counts per kg, or 2.33 mg per count. Those counts are not all useful: noise is far larger than one code.
With the datasheet’s typical 50 nV RMS at gain 128 and 10 SPS, peak-to-peak noise is about 330 nV. On a 10 kg, 2 mV/V cell at 5 V, that is about 330 mg before averaging, or 594 mg at 80 SPS. Real boards with long leads, shared supplies and vibration do worse.
Because a 24-bit code is about 2.33 nV at gain 128 and the input noise is many times that. Noise of 50 nV RMS is 21.5 counts RMS, so the bottom 7.1 bits of each reading move at random. Averaging or a slower data rate reduces it.
In counts per kilogram, no, if the ADC reference follows the excitation: the signal and the reference scale together. In noise-free resolution, yes, because the signal grows while the input-referred noise stays about the same. Never exceed the cell’s rated excitation, and for an ADC limited to 5 V, divide the reference down if the bridge runs higher.
Yes for both. At 3.3 V the HX711 accepts inputs from 1.2 V to 2.0 V (AGND + 1.2 V to AVDD − 1.3 V) and the ADS1232 at gain 64 or 128 accepts 1.5 V to 1.8 V (AGND + 1.5 V to AVDD − 1.5 V), and a bridge excited from that supply sits at 1.65 V. The ADS1232 needs at least 3 V for this to work at those gains. The bridge signal falls with excitation, so the noise-free resolution is worse than at 5 V.