How do you convert thermocouple voltage to temperature?
Add the voltage for the cold-junction temperature to the measured voltage, then look up the sum in the NIST reference function for that type. A type K reading 3.096 mV with terminals at 25 °C has a cold-junction voltage of 1.000 mV, so the sum is 4.096 mV, which is 100.0 °C.
Thermocouple formula: NIST reference functions and cold junction
NIST defines each type with polynomial reference functions on the ITS-90 scale, in millivolts with the reference junction at 0 °C.
E = Σ c_i t^i
- E
- thermoelectric voltage in mV, referenced to 0 °C
- t
- temperature in °C
- c_i
- published coefficients for the type and sub-range
Type K above 0 °C has an extra exponential term, which NIST defines with three more coefficients:
E = Σ c_i t^i + a0 exp(a1 (t − a2)²)
A meter connected at a terminal temperature Tcj reads the difference between the two ends:
V = E(T) − E(Tcj)
- V
- voltage at the terminals in mV
- T
- hot-junction temperature in °C
- Tcj
- cold-junction (terminal) temperature in °C
To go from thermocouple mV to temperature, add the cold-junction voltage first, then convert the sum:
The Seebeck coefficient is the slope of the reference function, dE/dt. NIST also publishes inverse polynomials of the form t = Σ d_i E^i, which are approximations. The calculator solves the reference function directly and lists the inverse polynomial result beside it.
How to calculate thermocouple voltage and temperature
Example 1: type K at 100 °C with terminals at 25 °C
E(100 °C) = 4.096 mV and E(25 °C) = 1.000 mV, so the meter reads 4.096 − 1.000 = 3.096 mV.
Example 2: type J reading 15 mV with terminals at 30 °C
E(30 °C) = 1.537 mV, so the compensated voltage is 15 + 1.537 = 16.537 mV. The reference function gives 303.78 °C, and the NIST inverse polynomial gives 303.78 °C.
Example 3: what skipping compensation costs
A type K reads 10 mV with terminals at 30 °C. Converting 10 mV with no compensation gives 246.23 °C. Adding the terminal temperature gives 276.23 °C. Adding the voltages first, which is correct, gives 275.65 °C, so the shortcut is off by 0.6 °C.
Thermocouple chart: voltage table in mV for types K, J, T, E and N
EMF in millivolts with the reference junction at 0 °C, every 50 °C, from the NIST reference functions. A dash means the type is not defined at that temperature.
| °C | Type K | Type J | Type T | Type E | Type N |
|---|
| −250 | −6.404 | — | −6.180 | −9.718 | −4.313 |
| −200 | −5.891 | −7.890 | −5.603 | −8.825 | −3.990 |
| −150 | −4.913 | −6.500 | −4.648 | −7.279 | −3.336 |
| −100 | −3.554 | −4.633 | −3.379 | −5.237 | −2.407 |
| −50 | −1.889 | −2.431 | −1.819 | −2.787 | −1.269 |
| 0 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 |
| 50 | 2.023 | 2.585 | 2.036 | 3.048 | 1.340 |
| 100 | 4.096 | 5.269 | 4.279 | 6.319 | 2.774 |
| 150 | 6.138 | 8.010 | 6.704 | 9.789 | 4.302 |
| 200 | 8.138 | 10.779 | 9.288 | 13.421 | 5.913 |
| 250 | 10.153 | 13.555 | 12.013 | 17.181 | 7.597 |
| 300 | 12.209 | 16.327 | 14.862 | 21.036 | 9.341 |
| 350 | 14.293 | 19.090 | 17.819 | 24.964 | 11.136 |
| 400 | 16.397 | 21.848 | 20.872 | 28.946 | 12.974 |
| 450 | 18.516 | 24.610 | — | 32.965 | 14.846 |
| 500 | 20.644 | 27.393 | — | 37.005 | 16.748 |
| 550 | 22.776 | 30.216 | — | 41.053 | 18.672 |
| 600 | 24.905 | 33.102 | — | 45.093 | 20.613 |
| 650 | 27.025 | 36.071 | — | 49.116 | 22.566 |
| 700 | 29.129 | 39.132 | — | 53.112 | 24.527 |
| 750 | 31.213 | 42.281 | — | 57.080 | 26.491 |
| 800 | 33.275 | 45.494 | — | 61.017 | 28.455 |
| 850 | 35.313 | 48.715 | — | 64.922 | 30.416 |
| 900 | 37.326 | 51.877 | — | 68.787 | 32.371 |
| 950 | 39.314 | 54.956 | — | 72.603 | 34.319 |
| 1000 | 41.276 | 57.953 | — | 76.373 | 36.256 |
| 1050 | 43.211 | 60.890 | — | — | 38.179 |
| 1100 | 45.119 | 63.792 | — | — | 40.087 |
| 1150 | 46.995 | 66.679 | — | — | 41.976 |
| 1200 | 48.838 | 69.553 | — | — | 43.846 |
| 1250 | 50.644 | — | — | — | 45.694 |
| 1300 | 52.410 | — | — | — | 47.513 |
| 1350 | 54.138 | — | — | — | — |
Range and Seebeck coefficient by type
Seebeck coefficients in µV/°C, taken from the slope of each reference function.
| Type | Positive / negative | Range (°C) | EMF span (mV) | at −100 °C | at 0 °C | at 100 °C | at 500 °C | at 1000 °C |
|---|
| K | Chromel (Ni-Cr) / Alumel (Ni-Al) | −270 to 1372 | −6.458 to 54.886 | 30.5 | 39.5 | 41.4 | 42.6 | 39.0 |
| J | Iron / Constantan (Cu-Ni) | −210 to 1200 | −8.095 to 69.553 | 41.1 | 50.4 | 54.4 | 56.0 | 59.3 |
| T | Copper / Constantan (Cu-Ni) | −270 to 400 | −6.258 to 20.872 | 28.4 | 38.7 | 46.8 | — | — |
| E | Chromel (Ni-Cr) / Constantan (Cu-Ni) | −270 to 1000 | −9.835 to 76.373 | 45.2 | 58.7 | 67.5 | 80.9 | 75.2 |
| N | Nicrosil (Ni-Cr-Si) / Nisil (Ni-Si-Mg) | −270 to 1300 | −4.345 to 47.513 | 20.9 | 26.2 | 29.6 | 38.3 | 38.6 |
NIST inverse polynomial ranges and errors
The sub-ranges NIST publishes inverse polynomials for, with the error NIST states for each. Use these figures if you implement the polynomials in firmware.
| Type | Temperature (°C) | Voltage (mV) | Stated error (°C) |
|---|
| K | −200 to 0 | −5.891 to 0.000 | −0.02 to 0.04 |
| K | 0 to 500 | 0.000 to 20.644 | −0.05 to 0.04 |
| K | 500 to 1372 | 20.644 to 54.886 | −0.05 to 0.06 |
| J | −210 to 0 | −8.095 to 0.000 | −0.05 to 0.03 |
| J | 0 to 760 | 0.000 to 42.919 | −0.04 to 0.04 |
| J | 760 to 1200 | 42.919 to 69.553 | −0.04 to 0.03 |
| T | −200 to 0 | −5.603 to 0.000 | −0.02 to 0.04 |
| T | 0 to 400 | 0.000 to 20.872 | −0.03 to 0.03 |
| E | −200 to 0 | −8.825 to 0.000 | −0.01 to 0.03 |
| E | 0 to 1000 | 0.000 to 76.373 | −0.02 to 0.02 |
| N | −200 to 0 | −3.990 to 0.000 | −0.02 to 0.03 |
| N | 0 to 600 | 0.000 to 20.613 | −0.02 to 0.03 |
| N | 600 to 1300 | 20.613 to 47.513 | −0.04 to 0.02 |
Standard wire tolerance
Types J, K, T and E are the IEC-EN 60584-2 limits as listed in Texas Instruments SBAA274A. Type N is the IEC 60584-1 table as reproduced by TC Direct and WIKA. The tolerance is the larger of the two terms.
| Type | Class | Range (°C) | Fixed term | Proportional term |
|---|
| J | Class 1 | −40 to 750 | ±1.5 °C | ±0.004 × |t| |
| J | Class 2 | −40 to 750 | ±2.5 °C | ±0.0075 × |t| |
| K | Class 1 | −40 to 1000 | ±1.5 °C | ±0.004 × |t| |
| K | Class 2 | −40 to 1200 | ±2.5 °C | ±0.0075 × |t| |
| K | Class 3 | −200 to 40 | ±2.5 °C | ±0.015 × |t| |
| T | Class 1 | −40 to 350 | ±0.5 °C | ±0.004 × |t| |
| T | Class 2 | −40 to 350 | ±1 °C | ±0.0075 × |t| |
| T | Class 3 | −200 to 40 | ±1 °C | ±0.015 × |t| |
| E | Class 1 | −40 to 800 | ±1.5 °C | ±0.004 × |t| |
| E | Class 2 | −40 to 900 | ±2.5 °C | ±0.0075 × |t| |
| E | Class 3 | −200 to 40 | ±2.5 °C | ±0.015 × |t| |
| N | Class 1 | −40 to 1000 | ±1.5 °C | ±0.004 × |t| |
| N | Class 2 | −40 to 1200 | ±2.5 °C | ±0.0075 × |t| |
| N | Class 3 | −200 to 40 | ±2.5 °C | ±0.015 × |t| |
This page treats each range as including its end points. TI prints the limits as strict inequalities, but WIKA tabulates types K and N at the ends: ±4.0 °C for class 1 at 1000 °C and ±9.00 °C for class 2 at 1200 °C. IEC 60584-1 is a paid document we did not read directly, so check it if a reading sits exactly on a limit.
What goes wrong in real circuits
The cold junction has to be measured where the wires land
Both terminals must sit at the same known temperature, which is why instruments use an isothermal block or a copper pour around the connector. A sensor a few centimeters away from a warm regulator reads a different temperature. At a 25 °C terminal and a 500 °C hot junction, each 1 °C of cold-junction error moves a type K reading by 0.95 °C.
The signal is small
At room temperature, type K produces about 40.5 µV/°C, so 10 µV of amplifier offset or noise is 0.25 °C. Any junction between dissimilar metals in the signal path forms another thermocouple, so keep the two input paths symmetric and at the same temperature.
Limits of this model
- The functions describe standard wire. A real thermocouple differs by up to its tolerance class, which is usually the largest error in the system.
- At the bottom of a type’s range the curve flattens, so microvolts correspond to large temperature changes.
- NIST gives no inverse polynomial below −200 °C for K, T, E and N.
- Extension wire and connectors must match the thermocouple type, or they add their own error.
Glossary
- Hot junction
- The welded tip of the thermocouple, at the temperature being measured.
- Cold junction
- The point where the thermocouple wires meet the measuring circuit, also called the reference junction.
- Cold-junction compensation
- Measuring the terminal temperature and adding its voltage so the result reads as if referenced to 0 °C.
- Seebeck coefficient
- The slope of the voltage curve, in µV/°C, which varies with temperature.
- ITS-90
- The International Temperature Scale of 1990, which the NIST reference functions are based on.
- Isothermal block
- A thermally massive connector that holds both thermocouple terminals at the same temperature.
Frequently asked questions
What is the thermocouple formula?
NIST defines each type with a polynomial reference function, E = Σ c_i t^i, in millivolts with the reference junction at 0 °C (type K adds an exponential term above 0 °C). A meter with its terminals at Tcj reads V = E(T) − E(Tcj), so add E(Tcj) back before converting to temperature.
How do I read a type K thermocouple table?
Find the temperature row and read the Type K column: 4.096 mV at 100 °C, with the reference junction at 0 °C. A type J thermocouple chart reads the same way. For a meter with warmer terminals, subtract the voltage at the terminal temperature, or enter both temperatures in the calculator above.
What voltage does a type K thermocouple produce at 100 °C?
4.096 mV with the reference junction at 0 °C. With the terminals at 25 °C the meter sees 3.096 mV, because the 1.000 mV of the cold junction is subtracted.
Why does a thermocouple need cold-junction compensation?
A thermocouple produces a voltage that depends on the difference between its two ends, and the NIST tables assume the reference end is at 0 °C. Measuring the terminal temperature and adding its voltage gives the value the tables expect.
Can I just add the cold-junction temperature to the reading?
Only as an approximation, because the curve is not straight. A type K reading 10 mV with terminals at 30 °C is 275.65 °C when compensated properly, while adding the temperatures gives 276.23 °C. The right method adds voltages, then converts.
What is the Seebeck coefficient of a type K thermocouple?
39.45 µV/°C at 0 °C, from the slope of the NIST reference function. It changes with temperature, so type K is about 42.6 µV/°C at 500 °C and 15.3 µV/°C at −200 °C.
What is the difference between type K, J, T, E and N?
The wire materials, range and sensitivity. Type K runs to 1372 °C, J to 1200 °C, T to 400 °C, E to 1000 °C and N to 1300 °C. Type E has the highest output of the five at 0 °C (58.7 µV/°C) and type N the lowest (26.2 µV/°C).
How accurate is a type K thermocouple?
The wire itself is the limit. IEC-EN 60584-2 class 2 allows ±2.5 °C or ±0.75 % of the temperature, whichever is larger, so ±3.75 °C at 500 °C. Class 1 allows ±1.5 °C or ±0.4 %.
Are the NIST inverse polynomials exact?
No. NIST publishes them as approximations, with stated errors of up to about ±0.06 °C, and only down to −200 °C (−210 °C for type J). This thermocouple calculator solves the reference function directly and shows the polynomial result beside it.
More free PCB tools
See all free electronics tools →
- Strain gauge calculatorBridge output for quarter, half and full strain gauge bridges from strain, gauge factor and excitation, or strain from output.
- Pt100 chart and RTD calculatorPT100 and PT1000 resistance at any temperature and temperature from resistance with the IEC 60751 Callendar-Van Dusen equation.
- Wheatstone bridge calculatorBridge output voltage, the balance condition and the unknown resistor of a Wheatstone bridge.
- NTC thermistor accuracy calculatorFull error budget for an NTC divider read by an ADC: thermistor, resistor, reference, ADC and self-heating errors across temperature, plus a firmware table.
- Transimpedance amplifier calculatorFeedback resistor and capacitor for a photodiode TIA, its bandwidth and stability, from the op amp GBW and input capacitance.
- Steinhart-Hart equation calculatorSteinhart-Hart A, B and C coefficients from three resistance-temperature points, Beta from two, and conversions with both.
Design the whole board, not one number
ProtoFlow is free AI PCB design software: describe a circuit and get a schematic with real parts, PCB layout with an autorouter, SPICE simulation and Gerber files. Native KiCad, Altium and EAGLE import and export. Free on macOS, Windows and Linux.