Thermocouple chart and calculator

Look up a voltage in the thermocouple chart below, or convert temperature to voltage and voltage to temperature for type K, J, T, E and N thermocouples with the NIST ITS-90 reference functions. Enter the terminal temperature and this thermocouple voltage calculator applies cold-junction compensation and shows the Seebeck coefficient at your operating point.

Convert
Examples
°C
°C

The temperature of the terminals. Use 0 for an ice-bath reference.

Voltage at the terminals

3.096 mV

100.0 °C hot, 25.0 °C cold

EMF of the hot junction E(T)
4.096 mV table value, 0 °C reference
EMF of the cold junction E(Tcj)
1.000 mV
Seebeck coefficient at the hot junction
41.37 µV/°C
Seebeck coefficient at the cold junction
40.52 µV/°C
Reading shift per 1 °C of cold-junction error
0.979 °C
Reading shift per 10 µV of input error
0.242 °C
Type K: EMF (0 °C reference) and Seebeck coefficient
-10010203040506001020304050-400-2000200400600800100012001400Temperature (°C)EMF (mV)Seebeck coefficient (µV/°C)100.0 °C
EMF (mV)Seebeck (µV/°C)

IEC 60584 tolerance of type K wire at 100.0 °C

ClassToleranceDefined overApplies here
Class 1±1.50 °C-40 to 1000 °CYes
Class 2±2.50 °C-40 to 1200 °CYes
Class 3±2.50 °C-200 to 40 °CNo

How do you convert thermocouple voltage to temperature?

Add the voltage for the cold-junction temperature to the measured voltage, then look up the sum in the NIST reference function for that type. A type K reading 3.096 mV with terminals at 25 °C has a cold-junction voltage of 1.000 mV, so the sum is 4.096 mV, which is 100.0 °C.

Thermocouple formula: NIST reference functions and cold junction

NIST defines each type with polynomial reference functions on the ITS-90 scale, in millivolts with the reference junction at 0 °C.

E = Σ c_i t^i
E
thermoelectric voltage in mV, referenced to 0 °C
t
temperature in °C
c_i
published coefficients for the type and sub-range

Type K above 0 °C has an extra exponential term, which NIST defines with three more coefficients:

E = Σ c_i t^i + a0 exp(a1 (t − a2)²)

A meter connected at a terminal temperature Tcj reads the difference between the two ends:

V = E(T) − E(Tcj)
V
voltage at the terminals in mV
T
hot-junction temperature in °C
Tcj
cold-junction (terminal) temperature in °C

To go from thermocouple mV to temperature, add the cold-junction voltage first, then convert the sum:

E(T) = V + E(Tcj)

The Seebeck coefficient is the slope of the reference function, dE/dt. NIST also publishes inverse polynomials of the form t = Σ d_i E^i, which are approximations. The calculator solves the reference function directly and lists the inverse polynomial result beside it.

How to calculate thermocouple voltage and temperature

Example 1: type K at 100 °C with terminals at 25 °C

E(100 °C) = 4.096 mV and E(25 °C) = 1.000 mV, so the meter reads 4.096 − 1.000 = 3.096 mV.

Example 2: type J reading 15 mV with terminals at 30 °C

E(30 °C) = 1.537 mV, so the compensated voltage is 15 + 1.537 = 16.537 mV. The reference function gives 303.78 °C, and the NIST inverse polynomial gives 303.78 °C.

Example 3: what skipping compensation costs

A type K reads 10 mV with terminals at 30 °C. Converting 10 mV with no compensation gives 246.23 °C. Adding the terminal temperature gives 276.23 °C. Adding the voltages first, which is correct, gives 275.65 °C, so the shortcut is off by 0.6 °C.

Thermocouple chart: voltage table in mV for types K, J, T, E and N

EMF in millivolts with the reference junction at 0 °C, every 50 °C, from the NIST reference functions. A dash means the type is not defined at that temperature.

°CType KType JType TType EType N
−250−6.404—−6.180−9.718−4.313
−200−5.891−7.890−5.603−8.825−3.990
−150−4.913−6.500−4.648−7.279−3.336
−100−3.554−4.633−3.379−5.237−2.407
−50−1.889−2.431−1.819−2.787−1.269
00.0000.0000.0000.0000.000
502.0232.5852.0363.0481.340
1004.0965.2694.2796.3192.774
1506.1388.0106.7049.7894.302
2008.13810.7799.28813.4215.913
25010.15313.55512.01317.1817.597
30012.20916.32714.86221.0369.341
35014.29319.09017.81924.96411.136
40016.39721.84820.87228.94612.974
45018.51624.610—32.96514.846
50020.64427.393—37.00516.748
55022.77630.216—41.05318.672
60024.90533.102—45.09320.613
65027.02536.071—49.11622.566
70029.12939.132—53.11224.527
75031.21342.281—57.08026.491
80033.27545.494—61.01728.455
85035.31348.715—64.92230.416
90037.32651.877—68.78732.371
95039.31454.956—72.60334.319
100041.27657.953—76.37336.256
105043.21160.890——38.179
110045.11963.792——40.087
115046.99566.679——41.976
120048.83869.553——43.846
125050.644———45.694
130052.410———47.513
135054.138————

Range and Seebeck coefficient by type

Seebeck coefficients in µV/°C, taken from the slope of each reference function.

TypePositive / negativeRange (°C)EMF span (mV)at −100 °Cat 0 °Cat 100 °Cat 500 °Cat 1000 °C
KChromel (Ni-Cr) / Alumel (Ni-Al)−270 to 1372−6.458 to 54.88630.539.541.442.639.0
JIron / Constantan (Cu-Ni)−210 to 1200−8.095 to 69.55341.150.454.456.059.3
TCopper / Constantan (Cu-Ni)−270 to 400−6.258 to 20.87228.438.746.8——
EChromel (Ni-Cr) / Constantan (Cu-Ni)−270 to 1000−9.835 to 76.37345.258.767.580.975.2
NNicrosil (Ni-Cr-Si) / Nisil (Ni-Si-Mg)−270 to 1300−4.345 to 47.51320.926.229.638.338.6

NIST inverse polynomial ranges and errors

The sub-ranges NIST publishes inverse polynomials for, with the error NIST states for each. Use these figures if you implement the polynomials in firmware.

TypeTemperature (°C)Voltage (mV)Stated error (°C)
K−200 to 0−5.891 to 0.000−0.02 to 0.04
K0 to 5000.000 to 20.644−0.05 to 0.04
K500 to 137220.644 to 54.886−0.05 to 0.06
J−210 to 0−8.095 to 0.000−0.05 to 0.03
J0 to 7600.000 to 42.919−0.04 to 0.04
J760 to 120042.919 to 69.553−0.04 to 0.03
T−200 to 0−5.603 to 0.000−0.02 to 0.04
T0 to 4000.000 to 20.872−0.03 to 0.03
E−200 to 0−8.825 to 0.000−0.01 to 0.03
E0 to 10000.000 to 76.373−0.02 to 0.02
N−200 to 0−3.990 to 0.000−0.02 to 0.03
N0 to 6000.000 to 20.613−0.02 to 0.03
N600 to 130020.613 to 47.513−0.04 to 0.02

Standard wire tolerance

Types J, K, T and E are the IEC-EN 60584-2 limits as listed in Texas Instruments SBAA274A. Type N is the IEC 60584-1 table as reproduced by TC Direct and WIKA. The tolerance is the larger of the two terms.

TypeClassRange (°C)Fixed termProportional term
JClass 1−40 to 750±1.5 °C±0.004 × |t|
JClass 2−40 to 750±2.5 °C±0.0075 × |t|
KClass 1−40 to 1000±1.5 °C±0.004 × |t|
KClass 2−40 to 1200±2.5 °C±0.0075 × |t|
KClass 3−200 to 40±2.5 °C±0.015 × |t|
TClass 1−40 to 350±0.5 °C±0.004 × |t|
TClass 2−40 to 350±1 °C±0.0075 × |t|
TClass 3−200 to 40±1 °C±0.015 × |t|
EClass 1−40 to 800±1.5 °C±0.004 × |t|
EClass 2−40 to 900±2.5 °C±0.0075 × |t|
EClass 3−200 to 40±2.5 °C±0.015 × |t|
NClass 1−40 to 1000±1.5 °C±0.004 × |t|
NClass 2−40 to 1200±2.5 °C±0.0075 × |t|
NClass 3−200 to 40±2.5 °C±0.015 × |t|

This page treats each range as including its end points. TI prints the limits as strict inequalities, but WIKA tabulates types K and N at the ends: ±4.0 °C for class 1 at 1000 °C and ±9.00 °C for class 2 at 1200 °C. IEC 60584-1 is a paid document we did not read directly, so check it if a reading sits exactly on a limit.

What goes wrong in real circuits

The cold junction has to be measured where the wires land

Both terminals must sit at the same known temperature, which is why instruments use an isothermal block or a copper pour around the connector. A sensor a few centimeters away from a warm regulator reads a different temperature. At a 25 °C terminal and a 500 °C hot junction, each 1 °C of cold-junction error moves a type K reading by 0.95 °C.

The signal is small

At room temperature, type K produces about 40.5 µV/°C, so 10 µV of amplifier offset or noise is 0.25 °C. Any junction between dissimilar metals in the signal path forms another thermocouple, so keep the two input paths symmetric and at the same temperature.

Limits of this model

  • The functions describe standard wire. A real thermocouple differs by up to its tolerance class, which is usually the largest error in the system.
  • At the bottom of a type’s range the curve flattens, so microvolts correspond to large temperature changes.
  • NIST gives no inverse polynomial below −200 °C for K, T, E and N.
  • Extension wire and connectors must match the thermocouple type, or they add their own error.

Glossary

Hot junction
The welded tip of the thermocouple, at the temperature being measured.
Cold junction
The point where the thermocouple wires meet the measuring circuit, also called the reference junction.
Cold-junction compensation
Measuring the terminal temperature and adding its voltage so the result reads as if referenced to 0 °C.
Seebeck coefficient
The slope of the voltage curve, in µV/°C, which varies with temperature.
ITS-90
The International Temperature Scale of 1990, which the NIST reference functions are based on.
Isothermal block
A thermally massive connector that holds both thermocouple terminals at the same temperature.

Related guides

Frequently asked questions

What is the thermocouple formula?

NIST defines each type with a polynomial reference function, E = Σ c_i t^i, in millivolts with the reference junction at 0 °C (type K adds an exponential term above 0 °C). A meter with its terminals at Tcj reads V = E(T) − E(Tcj), so add E(Tcj) back before converting to temperature.

How do I read a type K thermocouple table?

Find the temperature row and read the Type K column: 4.096 mV at 100 °C, with the reference junction at 0 °C. A type J thermocouple chart reads the same way. For a meter with warmer terminals, subtract the voltage at the terminal temperature, or enter both temperatures in the calculator above.

What voltage does a type K thermocouple produce at 100 °C?

4.096 mV with the reference junction at 0 °C. With the terminals at 25 °C the meter sees 3.096 mV, because the 1.000 mV of the cold junction is subtracted.

Why does a thermocouple need cold-junction compensation?

A thermocouple produces a voltage that depends on the difference between its two ends, and the NIST tables assume the reference end is at 0 °C. Measuring the terminal temperature and adding its voltage gives the value the tables expect.

Can I just add the cold-junction temperature to the reading?

Only as an approximation, because the curve is not straight. A type K reading 10 mV with terminals at 30 °C is 275.65 °C when compensated properly, while adding the temperatures gives 276.23 °C. The right method adds voltages, then converts.

What is the Seebeck coefficient of a type K thermocouple?

39.45 µV/°C at 0 °C, from the slope of the NIST reference function. It changes with temperature, so type K is about 42.6 µV/°C at 500 °C and 15.3 µV/°C at −200 °C.

What is the difference between type K, J, T, E and N?

The wire materials, range and sensitivity. Type K runs to 1372 °C, J to 1200 °C, T to 400 °C, E to 1000 °C and N to 1300 °C. Type E has the highest output of the five at 0 °C (58.7 µV/°C) and type N the lowest (26.2 µV/°C).

How accurate is a type K thermocouple?

The wire itself is the limit. IEC-EN 60584-2 class 2 allows ±2.5 °C or ±0.75 % of the temperature, whichever is larger, so ±3.75 °C at 500 °C. Class 1 allows ±1.5 °C or ±0.4 %.

Are the NIST inverse polynomials exact?

No. NIST publishes them as approximations, with stated errors of up to about ±0.06 °C, and only down to −200 °C (−210 °C for type J). This thermocouple calculator solves the reference function directly and shows the polynomial result beside it.